Parameter's name and its dimension Parameter's value
Spectral range (defined by image converter tube spectral sensitivity range), nm ≤400 - ≥800
Reverse linear dispersion, nm/mm 2.37
Spectral resolution, nm 0.08
Wavelength re-tuning step, nm 0.01
Frame size on image converter tube cathode, mm, not less than:
CCD readout unit X-axis is vertical
CCD readout unit X-axis is horizontal
11.2 ×16.7
12.5 ×11.2
Frame duration:
from, ns
up to, μs
8-12
600-660
Maximum slit image length on image converter tube cathode, mm, not less than:
CCD readout unit X-axis is horizontal
CCD readout unit X-axis is vertical
11.2
16.7
Effective sweep route length on image converter tube screen, mm, not less than:
readout unit X-axis is horizontal
readout unit X-axis is vertical
18
13
Coefficient of linear sweep on image converter tube screen:
from, ns/cm
up to, μs/cm
1
300
Temporal resolution with image converter tube slit width 0.05mm, not worse (not more) than:
from, ns
up to, μs
20
6
Maximum frequency of triggering, Hz:
only when triggering and repoducing image
during automatic image processing
1.0
0.5
Input (triggering) synchronous pulse parameters:
polarity
amplitude, V, not more than
duration, ns, not less than 
positive
50
100
Time of entering the mode of operation, min, not more than:
mode of functioning
mode of measuring spatial (spectral) and temporal intervals
1
30
Overall dimensons, mm × mm × mm 550 ×800 ×260
Weight, kg, not more than 40
Power supply voltage from industrial mains with 50-60 Hz frequensy, V 100-230
Power consumed from power supplying mains, W, not more than 75

 